Skip to main navigation Skip to search Skip to main content

What are typical sources of error in rotational rheometry of polymer melts?

Research output: Contribution to journalArticlepeer-review

14 Scopus citations

Abstract

Rheometers have made giant leaps in terms of usability, sensitivity, and versatility. This leads to the illusion that a rheometer can be used as a fool-proof device for measuring rheological properties. The article will focus on typical problems that are encountered in rheological practice when measuring polymer melts. Emphasis is put on problems related to measurement artefacts stemming from the rheometer as well as from the material itself. Furthermore, possibilities to eliminate rheometer related artefacts mostly related to the following phenomena -geometry inertia, thermal expansion, torque resolution, and environmental control -will be discussed. The sample related artefacts vary significantly from sample to sample and include: thermal degradation, nonlinear shear deformation, centrifugal forces, slip and shear banding, as well as miscibility, orientation, and distribution of different phases, of which only the ones occurring in homogeneous polymer melts are discussed here.

Original languageEnglish
Pages (from-to)277-291
Number of pages15
JournalKorea Australia Rheology Journal
Volume26
Issue number3
DOIs
StatePublished - Aug 2014
Externally publishedYes

Keywords

  • Artefacts
  • Inertia
  • Linearity and stationarity
  • Nonlinear regime
  • Thermal degradation

ASJC Scopus subject areas

  • General Materials Science
  • Condensed Matter Physics

Fingerprint

Dive into the research topics of 'What are typical sources of error in rotational rheometry of polymer melts?'. Together they form a unique fingerprint.

Cite this