Skip to main navigation Skip to search Skip to main content

STABILIZED CERAMIC MEMBRANE ELECTRODES FOR THE MEASUREMENT OF pH AT ELEVATED TEMPERATURES.

  • T. Tsuruta*
  • , D. D. Macdonald
  • *Corresponding author for this work

Research output: Contribution to journal › Article › peer-review

6 Scopus citations

Abstract

The use of a ceramic (ZrO//2 plus 17%U//2O//3) membrane for the measurement of pH at elevated temperatures is described. The measurement system consiste consists of matched membrane and external pressure-balanced electrodes, both of which have their internal Ag/AgCl sensing elements at ambient temperature in order to avoid thermal degradation. The pH response of the system has been calibrated using B(OH)//3 plus KOH and H//2SO plus Na//2SO//4 test solutions whose pH values were determined at various temperatures between 25 degree and 275 degree C using an in situ hydrogen electrode. The ceramic membrane has been shown to exhibit a Nernstein pH response, particularly at elevated temperatures (T greater than 200 degree C), although an ″acid error″ was observed for weakly acidic environments.

Original languageEnglish
Pages (from-to)1202-1210, 1221-1225
JournalJournal of the Electrochemical Society
StatePublished - 1982

ASJC Scopus subject areas

  • Electronic, Optical and Magnetic Materials
  • Renewable Energy, Sustainability and the Environment
  • Surfaces, Coatings and Films
  • Electrochemistry
  • Materials Chemistry

Fingerprint

Dive into the research topics of 'STABILIZED CERAMIC MEMBRANE ELECTRODES FOR THE MEASUREMENT OF pH AT ELEVATED TEMPERATURES.'. Together they form a unique fingerprint.

Cite this