Abstract
The use of a ceramic (ZrO//2 plus 17%U//2O//3) membrane for the measurement of pH at elevated temperatures is described. The measurement system consiste consists of matched membrane and external pressure-balanced electrodes, both of which have their internal Ag/AgCl sensing elements at ambient temperature in order to avoid thermal degradation. The pH response of the system has been calibrated using B(OH)//3 plus KOH and H//2SO plus Na//2SO//4 test solutions whose pH values were determined at various temperatures between 25 degree and 275 degree C using an in situ hydrogen electrode. The ceramic membrane has been shown to exhibit a Nernstein pH response, particularly at elevated temperatures (T greater than 200 degree C), although an ″acid error″ was observed for weakly acidic environments.
| Original language | English |
|---|---|
| Pages (from-to) | 1202-1210, 1221-1225 |
| Journal | Journal of the Electrochemical Society |
| State | Published - 1982 |
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Renewable Energy, Sustainability and the Environment
- Surfaces, Coatings and Films
- Electrochemistry
- Materials Chemistry
Fingerprint
Dive into the research topics of 'STABILIZED CERAMIC MEMBRANE ELECTRODES FOR THE MEASUREMENT OF pH AT ELEVATED TEMPERATURES.'. Together they form a unique fingerprint.Cite this