Abstract
The response function of a 125 mm diameter NE102A organic scintillation detector has been measured over the 2.7-14.8 MeV neutron energy range. The detector response function was derived from the light output for monoenergetic neutrons and gamma rays. The light output of the detector for monoenergetic neutrons was measured by selecting narrowgates in the time-of-flight (TOF) spectrum for a 241Am-Be neutron source. In order to provide check points on the data, the detector light output was also measured for monoenergetic neutrons from the D(d, n) and T(d, n) reactions. The response function of the NE102A detector is in good agreement (within 1-5%) with the published data of Cecil et al. [Nucl. Instr. and Meth. 161 (1979) 439].
| Original language | English |
|---|---|
| Pages (from-to) | 514-519 |
| Number of pages | 6 |
| Journal | Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment |
| Volume | 345 |
| Issue number | 3 |
| DOIs | |
| State | Published - 1 Jul 1994 |
Bibliographical note
Funding Information:This work is part of a KFUPM/RI project at ERL supported by the Research Institute of King Fahd University of Petroleum and Minerals .
ASJC Scopus subject areas
- Nuclear and High Energy Physics
- Instrumentation