Pulse height improvement of a 250 mm diameter NE213 scintillator due to a lightguide

  • A. A. Sunaidi
  • , A. A. Naqvi*
  • , H. A. Al-Juwair
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

Abstract

A systematic study has been carried out to determine the performance improvement in pulse height of a 250 mm diameter NE213 scintillator due to the use of a lightguide (LG). In this study the energy resolution, time resolution, and pulse shape discrimination (PSD) properties of the scintillator were measured with and without the LG. The energy resolution was measured using γ-γ coincidence technique while its time resolution was measured using time- correlated γ-rays from 60Co source. PSD tests of the scintillator were carried out using a 241Am-Be neutron source. After incorporating the LG, an improvement of up to of 54% and 28% was observed in the detector energy and time resolution respectively. The PSD characteristic did show a slight degradation due to the LG. The measured energy and time resolution of the detector with the LG is in agreement with published data.

Original languageEnglish
Pages (from-to)165-174
Number of pages10
JournalArabian Journal for Science and Engineering
Volume23
Issue number2A
StatePublished - Jul 1998

ASJC Scopus subject areas

  • General

Fingerprint

Dive into the research topics of 'Pulse height improvement of a 250 mm diameter NE213 scintillator due to a lightguide'. Together they form a unique fingerprint.

Cite this