Abstract
Performance of a 241Am-Be neutron source-based and 2.8 MeV neutrons-based moisture measurement setups have been compared using Monte Carlo simulation. In the setup fast neutrons transmitted through the sample were detected by a fast neutron detector, which was placed behind a massive long double truncated collimator. The setup geometry was optimized to detect maximum effect of 1-7 wt.% moisture on the neutron intensity transmitted through the sample. The yield of neutrons transmitted through concrete, coal, wood and soil samples containing 1-7 wt.% moisture was calculated for 2.8 MeV neutrons and neutrons from an 241Am-Be source. The slopes of the fast neutron intensities transmitted through the samples vs. their moisture contents are very sensitive to the neutron energy and the sample composition. Higher slopes have been observed for the samples with larger bulk density. The slopes of fast neutron yield show dependence on the incident neutron energy. Larger slopes have been observed for neutrons with samller energy. Due to the overall large slopes of the transmitted intensity data of the samples for 2.8 MeV neutrons, it is expected to achieve better sensitivity in moisture measurements for a 2.8 MeV neutrons based moisture setup.
| Original language | English |
|---|---|
| Pages (from-to) | 131-137 |
| Number of pages | 7 |
| Journal | Journal of Radioanalytical and Nuclear Chemistry |
| Volume | 274 |
| Issue number | 1 |
| DOIs | |
| State | Published - Oct 2007 |
Bibliographical note
Funding Information:The authors acknowledge the support of King Fahd University of Petroleum and Minerals, Dhahran, Saudi Arabia and Girls Education College, Riyadh Girls Colleges, Riyadh, Saudi Arabia.
ASJC Scopus subject areas
- Analytical Chemistry
- Nuclear Energy and Engineering
- Radiology Nuclear Medicine and imaging
- Pollution
- Spectroscopy
- Public Health, Environmental and Occupational Health
- Health, Toxicology and Mutagenesis