On Line Determination of Perphenazine Using Flow Injection Analysis

  • S. M. Sultan*
  • , A. M. Abdennabi
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

7 Scopus citations

Abstract

An accurate specific flow injection spectrophotometric method for the determination of perphenazine in the range 50-250 ppm is introduced. In the method, 110 μl of the drug is injected through a stream of 0.2% (w/v) potassium dichromate in 0.25 M sulfuric acid flowing on line as a carrier stream. The drug is oxidized on the flow to the red monocation radical, the peak absorbance of which is monitored at 525 nm. A throughput of up to 300 samples per hour is attained. The mechanism of the reaction is suggested and the method is compared with the United States Pharmacopeia method.

Original languageEnglish
Pages (from-to)343-348
Number of pages6
JournalMicrochemical Journal
Volume48
Issue number3
DOIs
StatePublished - Dec 1993

ASJC Scopus subject areas

  • Analytical Chemistry
  • Spectroscopy

Fingerprint

Dive into the research topics of 'On Line Determination of Perphenazine Using Flow Injection Analysis'. Together they form a unique fingerprint.

Cite this