Skip to main navigation Skip to search Skip to main content

Functional diagnosis of coronary artery stenoses using pressure drop coefficient: A pilot study in humans

  • Kranthi K. Kolli
  • , Mohamed Effat
  • , Imran Arif
  • , Tarek Helmy
  • , Massoud Leesar
  • , Lloyd H. Back
  • , Rupak K. Banerjee

Research output: Chapter in Book/Report/Conference proceeding › Conference contribution › peer-review

Abstract

Optimization of cut-off values for diagnostic parameters in determining stenosis severity depends on coupling functional (hemodynamic) and anatomical data. In this study, we sought to investigate the functional diagnostic parameter (CDP; the ratio of trans-stenotic pressure drop to distal dynamic pressure) by correlating with FFR (ratio of distal to proximal pressure of a stenotic section) and CFR (ratio of hyperemic blood flow velocity to that of resting condition). A 0.014-inch dual sensor (pressure and Doppler velocity) guidewire was used in 9 patients to obtain pressure drop and velocity at hyperemia. Functional index CDP was assessed from measured hyperemic pressure drop and velocity. CDP correlated linearly and significantly with FFR and CFR. Indicating that, measurements of CDP can provide a better assessment of stenosis severity.

Original languageEnglish
Title of host publicationASME 2011 Summer Bioengineering Conference, SBC 2011
Pages477-478
Number of pages2
EditionPARTS A AND B
DOIs
StatePublished - 2011
Externally publishedYes
EventASME 2011 Summer Bioengineering Conference, SBC 2011 - Farmington, PA, United States
Duration: 22 Jun 2011 → 25 Jun 2011

Publication series

NameASME 2011 Summer Bioengineering Conference, SBC 2011
NumberPARTS A AND B

Conference

ConferenceASME 2011 Summer Bioengineering Conference, SBC 2011
Country/TerritoryUnited States
CityFarmington, PA
Period22/06/11 → 25/06/11

ASJC Scopus subject areas

  • Bioengineering

Cite this