Energy resolution of NE102A and NE230 scintillators for monoenergetic gamma rays

A. A. Naqvi*, F. Z. Khiari, A. Coban, Abdulaziz M. Al-Jalal

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

11 Scopus citations

Abstract

The energy resolution of 125 and 50 mm diameter cylindrical NE102A detectors and 25 mm × 19 mm (height × diameter) NE230(C6D6) deuterated scintillator has been measured for monoenergetic gamma rays with an energy of 0.5-1.3 MeV. Over this energy range, the energy resolution of 125 mm NE102A detector varies from 22.8 to 13.2% and from 19.6 to 11.5% for the 50 mm NE102A detector. The measured energy resolution of the NE230 scintillator varies from 12.9 to 7.0% over the same energy range. The energy resolution of NE102A detectors and C6D6 scintillator is in good agreement with the energy resolution of NE213 detectors with comparable size.

Original languageEnglish
Pages (from-to)223-225
Number of pages3
JournalNuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
Volume324
Issue number1-2
DOIs
StatePublished - 1 Jan 1993

Bibliographical note

Funding Information:
This work is part of KFUPM/RI Project ERL supported by the Research Institute of King Fahd University of Petroleum and Minerals .