Energy dispersive X-ray fluorescence analysis of chloride in reforming catalysts

R. P. Singh*, R. P. Khatri, E. R. Pambid, N. M. Abbas

*Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

1 Scopus citations

Abstract

A relatively less expensive Phillips PV 9500/SY 585 automatic energy dispersive X-ray fluorescence (EDXFR) spectrometer with rhodium target, was modified using a commercial kitchen aluminium wrapping foil filter, for the determination of chloride in bifunctional reforming catalysts. Quantitation of chloride was carried out by the Claisse-Quintin method with a catalyst sample of known chloride content as an external standard. The results of EDXRF chloride analysis of different catalyst samples, containing varying amounts of percent carbon and chloride, compared well (p>0.75) with those obtained by silver chloride precipitation method. The small values of percent relative standard deviations at 0.4 to 4.5% showed a good reproducibility of the method. With the total analysis time of a few minutes combined with no sample preparation, the present method seems to be superior as compared to presently employed methods of chloride analysis in reforming catalysts.

Original languageEnglish
Pages (from-to)499-504
Number of pages6
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume142
Issue number2
DOIs
StatePublished - Oct 1990

ASJC Scopus subject areas

  • Analytical Chemistry
  • Nuclear Energy and Engineering
  • Radiology Nuclear Medicine and imaging
  • Pollution
  • Spectroscopy
  • Public Health, Environmental and Occupational Health
  • Health, Toxicology and Mutagenesis

Fingerprint

Dive into the research topics of 'Energy dispersive X-ray fluorescence analysis of chloride in reforming catalysts'. Together they form a unique fingerprint.

Cite this