Abstract
We describe a simple and sensitive porous polypropylene membrane-protected micro-solid-phase extraction (μ-SPE) approach for the sample preparation and determination of carbamate pesticides in soil samples by high-performance liquid chromatography. The μ-SPE device consisted of C18 sorbent held within a porous polypropylene envelope. In order to achieve optimum performance, several extraction parameters were optimized. Under the most favorable conditions, the extraction efficiency of the μ-SPE was very high, with detection limits in the range of 0.01-0.40 ng g-1. This is more than two orders of magnitude lower than the limits obtained by the United States Environmental Protection Agency Methods 8321A and 8318. A linear relationship was obtained for each analyte in the range of 2 and 200 ng g-1. The relative standard deviation for the analysis of aged soil samples spiked at 5 ng g-1 was ≤11%. The reproducibility of separate μ-SPE device used for experiments was satisfactory (relative standard deviations ranged from 4 to 11%), indicating that the method is reliable for routine environmental analysis.
| Original language | English |
|---|---|
| Pages (from-to) | 211-216 |
| Number of pages | 6 |
| Journal | Journal of Chromatography A |
| Volume | 1216 |
| Issue number | 2 |
| DOIs | |
| State | Published - 9 Jan 2009 |
| Externally published | Yes |
Bibliographical note
Funding Information:The authors gratefully acknowledge the financial support of this research by the National University of Singapore.
Keywords
- Environmental analysis
- Pesticides
- Solid-phase extraction and membrane extraction
ASJC Scopus subject areas
- Analytical Chemistry
- Biochemistry
- Organic Chemistry