An241Am-Be source based thermal neutron activation analysis facility at KFUPM

  • A. Aksoy*
  • , M. N. Al-Haddad
  • , A. A. Naqvi
  • , K. Al-Nahdi
  • *Corresponding author for this work

Research output: Contribution to journalArticlepeer-review

9 Scopus citations

Abstract

An instrumental thermal neutron activation analysis facility based, on a 16 Ci241 Am-Be source, a high resolution γ-ray spectrometry setup and a PC-based data acquisition system at KFUPM is described. The thermal neutron flux distribution was determined from the induced activities of high purity indium foils. The absolute thermal neutron flux was calculated from the activities of bare and cadmium-covered gold foils at a position of 3 cm from the soource at which the flux reaches a maximum. The facility tests were carried out with the determination of manganese concentrations in six types of industrially important steel samples. The result of 1.33% manganese in SS-304 steel sample was in excellent agreement with the literature value. The method is nondestructive, economical and ideal for bulk analysis.

Original languageEnglish
Pages (from-to)131-140
Number of pages10
JournalJournal of Radioanalytical and Nuclear Chemistry
Volume181
Issue number1
DOIs
StatePublished - Jun 1994

ASJC Scopus subject areas

  • Analytical Chemistry
  • Nuclear Energy and Engineering
  • Radiology Nuclear Medicine and imaging
  • Pollution
  • Spectroscopy
  • Public Health, Environmental and Occupational Health
  • Health, Toxicology and Mutagenesis

Fingerprint

Dive into the research topics of 'An241Am-Be source based thermal neutron activation analysis facility at KFUPM'. Together they form a unique fingerprint.